MATERIALS
Materials Provided
Test devices Droppers Buffer Package insert
Materials Required But Not Provided
Specimen collection containers Lancets (for fingerstick whole blood only)
Centrifuge Micropipette Timer
DIRECTIONS FOR USE
Allow test cassette, specimen, buffer,and/or controls to equilibrate to room temperature (15-30°C) prior to
testing.
1. Remove the test cassette from the foil pouch and use it as soon as possible. Best results will be obtained if the
assay is performed within one hour.
2. Place the test cassette on a clean and level surface. Be sure to label the device with specimen’s ID number.
3. Dip the sampling loop into specimen totally and rotate 2-3 cycles. Holding the sampling loop vertically, put the
sampling loop into the sample well (S) till touching sample pad (approx.3μL). Then add 2 drops(approx. 50-70μL) of
buffer immediately into the buffer well (B) and start the timer. Avoid air bubbles. See illustration below.
4. Wait for the red line(s) to appear. The result should be read in 15 minutes. Do not interpret the result after 15
minutes.